New Site, New Features!   Advanced search filters are here – Create a new account to get started!

  • Home
  • About Us
  • Resource Centre
  • News
1300 186 107
HOBO Data Loggers
  • Products
    • By Measurement
      • CO2
      • Temperature
      • Water Temperature
      • Water Level
      • Soil Moisture
      • Shop All
    • By Product Type
      • Data Loggers
      • Sensors
      • Kits & Systems
      • Software
      • Accessories
      • Shop All
    • By Environment
      • Climate Controlled
      • Indoor
      • Outdoor
      • Underwater
      • Shop All
    • Take your monitoring to the next level with HOBO's MX800 Multiparameter Water Quality Data Loggers
  • Applications
    • Building Performance
      • Air Quality
      • Energy/Power
      • Warehousing
    • Field Monitoring & Research
      • HOBOnet Wireless Networks
      • Soil Health
    • Water
      • Water Quality
      • Water Level
      • Water Temperature
      • Wetlands Monitoring
    • Outdoor
      • Agriculture
      • Irrigation
      • Weather
    • Pharmaceuticals
      • Fridge/Freezer Monitoring
      • Shipment & Storage Monitoring
  • Support
    • Resource Centre
      • Quick Start Guides
      • User Manuals
      • Datasheets
      • Webinars
      • Videos
    • Whitepaper Featured
      • Choosing a Water Level Logger: 5 Things you Should Know
    • Webinar Featured
      • HOBO Water Monitoring Platform for the Future Webinar
    • How-To Video Featured
      • Choosing a Remote Field Monitoring System – Factors to Consider
  • Home
  • About Us
  • Resource Centre
  • News
HOBO Data Loggers
  • Products
    • By Measurement
      • CO2
      • Temperature
      • Water Temperature
      • Water Level
      • Soil Moisture
      • Shop All
    • By Product Type
      • Data Loggers
      • Sensors
      • Kits & Systems
      • Software
      • Accessories
      • Shop All
    • By Environment
      • Climate Controlled
      • Indoor
      • Outdoor
      • Underwater
      • Shop All
    • Take your monitoring to the next level with HOBO's MX800 Multiparameter Water Quality Data Loggers
  • Applications
    • Building Performance
      • Air Quality
      • Energy/Power
      • Warehousing
    • Field Monitoring & Research
      • HOBOnet Wireless Networks
      • Soil Health
    • Water
      • Water Quality
      • Water Level
      • Water Temperature
      • Wetlands Monitoring
    • Outdoor
      • Agriculture
      • Irrigation
      • Weather
    • Pharmaceuticals
      • Fridge/Freezer Monitoring
      • Shipment & Storage Monitoring
  • Support
    • Resource Centre
      • Quick Start Guides
      • User Manuals
      • Datasheets
      • Webinars
      • Videos
    • Whitepaper Featured
      • Choosing a Water Level Logger: 5 Things you Should Know
    • Webinar Featured
      • HOBO Water Monitoring Platform for the Future Webinar
    • How-To Video Featured
      • Choosing a Remote Field Monitoring System – Factors to Consider
    • My account
    • Quote
    • Cart
←
→
  • OVERVIEW

    The pHionics STs Series™ Dissolved Oxygen sensor accurately measures dissolved oxygen (0-20 mg/L or percent saturation) and temperature (0-50°C) over long periods of time with little maintenance required. It outputs two noise-resistant 4-20 mA signals compatible with most dataloggers or remote telemetry units. A narrow diameter allows the STs Series to go where other sensors cannot while a durable, chemical-resistant housing ensures many years in-use. The electrode tip (anode) has a large surface area that improves oxygen exchange consistency across the membrane while reducing the frequency of fouling.

    In contrast to other dissolved oxygen sensors that require membrane and reference solution replacement every 3-6 months, the STs Series Dissolved Oxygen typically only requires electrode replacement every 2-3 years. This is possible because the electrode is permanently sealed to maintain a stable and sterile internal environment. In addition, the rugged Teflon® membrane resists scratches and tears that would put other sensors out of commission. All these features free up the user to focus on actionable insights from the data instead of spending time performing maintenance.

    Ranges available:

    0-20 mg/L or 0-200% saturation depending on scaling with datalogger or RTU

    STs Series Dissolved Oxygen Sensor Features List

    • Galvanic electrode with a large surface area anode to increase stability and prevent clogging
    • Secondary 0-50°C temperature sensor
    • Chemical-resistant Delrin®, PVDF, and 316 Stainless Steel housing (Titanium optional) for long-term monitoring
    • Narrow diameter (0.75 inches or 1.9 cm)
    • Removable electrode guard to reduce debris buildup and protect from damage
    • Fully submersible up to 70 psi
    • Shielded, Kevlar®-reinforced, water-blocked polyurethane cable
    • 4-20 mA signal capable of transmission up to 3 miles (with proper wire gauge)
    • Patented pHiConn™ connections to easily replace any part without special tools
    • Automatic Temperature Compensation (2% per degree Celsius)
    • Reversible Input Protection and Operation
    • Low power usage for long-term water quality monitoring
    • Isolated to prevent ground loops
    • Signal conditioning to remove background noise

    Accurate Measurements

    Application List

    • Industrial Effluent Monitoring
    • Wastewater Treatment
    • Environmental Monitoring
    • Aquaculture
  • SPECIFICATIONS

    Dissolved Oxygen Sensor

    Output 4-20mA
    Dissolved Oxygen Range 0-20 mg/L (ppm)
    Electrode Type Galvanic cell with Teflon membrane
    Minimum Flow Rate 1..5 ft (0.457m) per second
    Transmitter Type 2-wire
    Automatic Temperature Compensation 2% per °C
    Isolation 600 VDC
    Operational Temperature Range 0-70°C
    Power Supply Voltage 8-40 VDC
    Loop Impedance (Max) 200 Ohms at 12 VDC

    800 Ohms at 24 VDC

    1600 Ohms at 40 VDC

    Response time 98% < 60 seconds
    Pressure Range 0-70psi
    Wetted Materials 316 Stainless Steel, Delrin®, Viton™, Teflon®, epoxy
    Length 14.25in (36.20cm)
    Width 0.750 in (1.9cm) maximum
    Weight (No scale) 0.50lbs (0.227 kg)
    Linearity ± 0.5% of full scale*
    Accuracy ± 2.0% of full scale*
    Sensitivity ± 0.05% of full scale*
    Stability ± 2.0% of full scale*
    Repeatability ± 1.0% of full scale*

    Temperature Sensor

    Output 4-20mA
    Temperature Range 0-50°C
    Accuracy +/- 1°C
    Linearity +/- 0.5°C
    Power Supply Voltage 7-40 VDC

    Cable

    Cable Material Polyurethane
    Type 4 conductor, Kevlar®, reinforced, water blocked, shielded, twisted-pair wiring
    Maximum Distance 3 miles
    Resistance per 1000ft (304.8m) 26 Ohms
    Replaceable? Yes

    *After proper calibration

  • FAQ

    How long does a new electrode take to burn off oxygen in the cell?

    New dissolved oxygen electrodes have trapped oxygen from the manufacturing process.  This trapped oxygen must be burned off before accurate measurement can be taken.  While many dissolved oxygen electrodes require multiple days, our electrodes burn off trapped oxygen within 15 minutes.

    How should the electrode be stored?

    Please see the STs Series Storage guide >>

    How long will the electrode last?

    The standard life is 24-30 months while engaged with the sensor, powered or not. To extend the life, disengage the electrode from the transmitter at least one full turn, or completely remove the electrode and store it.  Always protect the electrode with the provided rubber boot or similar. Unlike many dissolved oxygen electrodes, ours do not have any internal resistors or connections so they do not passively consume the electrolytes until attached to the sensor.

    Do you provide a warranty?

    Yes, our products come with a 2-year warranty, excluding the electrode. These products are designed to last far longer if properly cared for and we will do what we can to help if any problems occur in or out of warranty.  Please don’t hesitate to reach out!

    Read our full warranty here >>

    What voltage is required for the STs Series Dissolved Oxygen?

    An 8VDC power supply is sufficient with standard lengths of cable.  Voltage drop must be accounted for over long lengths of cable (50+ ft.) to ensure enough power is supplied to any water quality sensor.  To calculate it, first double the length of cable to get the account for the full length of the circuit, then multiply that by resistance per foot/meter to get total resistance.  pHionics uses 24 AWG tinned-copper wire which has a resistance of 0.026 ohms per foot (0.084 ohms per meter).  Multiply total resistance by the maximum expected current (20 mA, or 0.02 A, is safest) to obtain expected voltage drop.  Verify that the voltage drop plus required voltage for pHionics sensors (8 VDC) is less than the power supply voltage.

    How long are STs Series devices expected to last?

    While the electrode will require replacement every two years on average, the housing, seals, and cables are intended to last 10-20 years with proper care. To increase longevity of the seals, the Viton O-rings should be relubricated with a silicon grease during each disassembly, or at least inspected for any debris or damage.  The harshness of the solution (high/low pH, extreme salinity, high pressure, etc.) also affects longevity.  All parts are replaceable to save money and limit waste in the event that any piece gets damaged.

    What type of electrode is provided?

    Our dissolved oxygen electrode is a galvanic cell containing propionic acid.  It has a larger surface area than other electrodes, allowing for extremely fast response time and a longer lifespan.

    What spare parts are recommended?

    For critical measurements, a duplicate STs Series sensor is always recommended so a replacement for every part is available at any time.  A duplicate simplifies troubleshooting by always having a functional STs Series available.  It also for swapping of devices during calibration and cleaning, so measurement is not significantly disrupted.  At a minimum, a spare electrode is recommended in the event of damage or eventual replacement.

    How often should the STs Dissolved Oxygen Series be calibrated and/or cleaned?

    The frequency of calibration must be determined empirically as rate of drift and debris accumulation differs for each application. Calculating the offset during calibration provides information on whether the electrode is still viable.  See our article on Proper Sensor Calibration for assistance in the process.  Check out our Dissolved Oxygen Electrode Care Guide for cleaning and troubleshooting procedures.

    Is the output automatically temperature-compensated (ATC)?

    The output is adjusted by 2% per °C relative to 25°C.  A thermistor within the housing is used which is why the ATC is most accurate when the device is fully submersed.  ATC functions regardless of whether the temperature output is in use.

    How can the STs Series be installed/mounted?

    The STs Series can be mounted in-line or immersed with the cmp-k compression fitting in a 3/4″ pipe, or freely suspended by the Kevlar™ cable.  Please see the pHionics STs Series Device Manual or contact us for more information.

/ Shop / Products / pHionics / pHionics STs - DO (Dissolved Oxygen) Sensor with 4-20mA

pHionics STs – DO (Dissolved Oxygen) Sensor with 4-20mA

$3,072.00 ex GST

The pHionics STs Series™ Dissolved Oxygen sensor accurately measures dissolved oxygen (0-20 mg/L or percent saturation) and temperature (0-50°C) over long periods of time with little maintenance required.

Part Number: STS3020S-025
Delivery Timeframe: Delivery Date: 2-3 weeks
Category: Products
Unit Price: $3,072.00

$3,072.00 ex GST

Add to quote loading

Product Resources

←
→

STs DO Sensor Quick Start Guide

D-pHi Series Device Manual

pHionics pH & Dissolved Oxygen Electrode Care and…

RX3000 Setup with pHionics STs Series Devices User…

pHionics STs Series Sensor Initial Setup How-to-Video

pHionics STs Series™ Setup with HOBO RX3000 How-to-Video

pHionics pHiKlip™ System for Multiparameter Water Quality Measurements

Troubleshooting pHionics STs Series Sensors How-to-Video

Proper Storage of pHionics pH Electrodes How-to-Video

pHionics STs Series Electrode Replacement How-to-Video

pHionics STs Series Cable Replacement How-to-Video

You May Also Like

←
→

HOBO® U20 Series Water Level Loggers

$1,390.00 – $1,715.00 ex GST

VIEW PRODUCT

HOBOnet MicroRX Station

Variable pricing

VIEW PRODUCT

HOBO Plug for Sensor Inputs | MX800 Series

$58.00 ex GST

VIEW PRODUCT

HOBOnet Wireless Temperature Sensor (Battery Powered)

$418.00 – $441.00 ex GST

VIEW PRODUCT

HOBO Split-core AC Current Sensors

$386.00 – $483.00 ex GST

VIEW PRODUCT

HOBO Smart Sensor Extension Cable

$78.00 – $186.00 ex GST

VIEW PRODUCT

HOBO MX800 Series Water Data Logger

Variable pricing

VIEW PRODUCT

HOBO Self-Describing DC Voltage Input Cable Sensor

$80.00 – $105.00 ex GST

VIEW PRODUCT

HOBO U30 - USB (Stand Alone) Data Logger

$1,507.00 – $1,808.00 ex GST

VIEW PRODUCT

HOBO 12-Bit Temperature Smart Sensors

$236.00 – $322.00 ex GST

VIEW PRODUCT

Fast Delivery: Free for orders over $100

Secure Online Payments: Visa or Mastercard

Quick Links

  • Products
  • Resource Centre
  • About HOBO Data Loggers
  • My Account
  • Contact Us

HEAD OFFICE

1300 186 107
[email protected]

Sydney  |  Melbourne  |  Brisbane  |  Perth  |  Adelaide

Quality & Security Guaranteed

  • Shipping & Returns
  • Terms and Conditions
  • Privacy Policy

Copyright © 2025 OneTemp. All rights reserved.